Fuse Replacement If the system fails to operate when plugged in and
switched on, the two fuses situated above the mains
power cord on the pump should be checked. To do this,
disconnect the pump from the main power supply and
remove each fuse holder cap using a suitable flat-bladed
screwdriver.
Caution
To protect against fire hazard, replace blown fuses only with the
identical type and rating (Refer to Section 10, Page 29 “Technical
Description”).
If either fuse blows again contact a service engineer or suitably
qualified personnel.