Sonicaid FM800 Series Reference Manual
100
15.3 Transducers
Ultrasound
Range
30 to 240 bpm
Accuracy
± 1 bpm over the range 100
-180 bpm
Alarms
High and Low FHR: 30 to 240 bpm
Signal Loss: % loss in last 5 minutes
Mode
Direction
al pulsed Doppler
Repetition rate 3.0KHz
Frequency
1.5MHz (yellow)
2.0MHz (blue)
P
<1Mpa
Iob
<20mW/cm
2
Ispta
<100mW/cm
2
Resolution
12 bits
Safety
Type CF protection
FECG
Range
30 to 240 bpm
Accuracy
± 1 bpm over the range 100
180 bpm
Alarms
As
Ultrasound
Common mode rejection
Better than 66db within the signal bandwidth
Input impedance
10M Ohm
Input range
>
30µV to >500µV peak to peak
DC offset
±2V common mode
±300mV differential
Common mode range
±20V @ mains frequency
Noise
<10µV peak
-to-peak referred to input
Safety
Type CF protection