For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer‘s declaration - electromagnetic immunity
The AD-2021 is intended for use in the electromagnetic environment speci ed below.
The customer or the user of the AD-2021 should assure that it is used in such an
environment.
IMMUNITY test IEC 60601 test
level
Compliance
level
Electromagnetic
environment-guidance
Radiated RF
IEC 61000-4-3
3 V/m 80 MHz
to 2.5 GHz
3 V/m Portable and mobile RF commu-
nications equipment should be
used no closer to any part of the
AD-2021, including cables, than
the recommended separation dis-
tance calculated from the equation
applicable to the frequency of the
transmitter. Recommended separa-
tion distance:
80 MHz to 800 MHz
800 MHz to 2,5 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the trans-
mitter manufacturer and d is the
recommended separation distance
in meters (m).
Field strengths from  xed RF
transmitters, as determined by an
electromagnetic site survey,
a
should
be less than the compliance level in
each frequency range.
b
Interference may occur in the
vicinity of equipment.
Marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation
is a ected by absorption and re ection from structures, objects and people.
a
: Field strengths from  xed transmitters, such as base stations for radio (cellular/cord-
less) telephones and land mobile radios, amateur radio, AM and FM radio broadcast
57
08428_DE-GB-FR-NL_V2.indd 5728_DE-GB-FR-NL_V2.indd 57 19.12.2013 10:18:5619.12.2013 10:18:56