For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration - electromagnetic emissions
The AD-2021 is intended for use in the electromagnetic environment speci ed below.
The customer or the user of the AD-2021 should assure that it is used in such an
environment.
Emissions test Compliance
Electromagnetic
environment-guidance
RF emissions
CISPR 11
Group 1
The AD-2021 uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electro-
nic equipment
RF emissions CISPR 11 Class B
The AD-2021 is suitable for use in all
establishments other than domestic and
those directly connected to the public
low-voltage power supply network that
supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not
applicable
Voltage  uctuations/
icker emissions IEC
61000-3-3
Not
applicable
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration - electromagnetic immunity
The AD-2021 is intended for use in the electromagnetic environment speci ed below.
The customer or the user of the AD-2021 should assure that it is used in such an
environment.
IMMUNITY test IEC 60601 test
level
Compliance
level
Electromagnetic
environment-guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile. If
oors are covered with
synthetic material, the relative
humidity should be at least
30 %.
Power frequency
(50/60 Hz)
magnetic  eld
3 A/m 3 A/m Power frequency magnetic
elds should be at levels
characteristic of a typical loca-
tion in a typical commercial or
hospital environment.
IEC 61000-4-8
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
ELECTROMAGNETIC COMPATIBILITY INFORMATION
56