Page 34
9.3 Removal and reinstallation of
the steam cylinder
Risk of eye injuries!
The clips that fix the steam cylinder halves
have sharp edges and can jump off during
dismantling.
Eye injuries are possible.
Wear proper PPE (Personal Protection
Equipment)!
Steam cylinder removal
» Remove unit housing cover
» Remove SuperFlush solenoid hose
from cylinder bottom (if applicable).
» Remove steam hose from steam
hose adapter.
If the steam hose is not to be disconnected,
the steam hose adapter with the steam hose
still attached may be detached from the
steamcylinder as shown in the next figure.
&$87,21
Set control
switch to „II“ po-
sition for residual
water draining
When cylinder is empty, set
control switch to „0“ position
Disconnect unit from power supply
Verify safe isolation
Current clamp
Disconnect from water
supply
Wait 10 mins. Then check cy-
linder temperature by cautious
approximation with hand (do
not touch!)