IACER Srl 27 MNPG69-09
Guidance and manufacturer's declaration ELECTROMAGNETIC IMMUNITY
FOR ALL DEVICES AND SYSTEMS
MAG2000 is intended for use in the electromagnetic environment specified
below. The user or operator of MAG2000 should assure that is used in such
environment.
Immunity test
Test level
IEC 60601
Compliance
level
Electromagnetic
environment
guidance
IEC 61000-4-4
commercial or hospital
environment.
Impulses
IEC 61000-4-5
±1kV
line - line
±1kV
line - line
Mains power quality
should be at least that
one of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dips of
U
T
)
for 0,5 cycle
<5% U
T
(>95% dips of
U
T
)
for 1 cycle
70% U
T
(30% dips of
U
T
)
for 25 cycles
<5% U
T
(>95% dips of
U
T
)
for 5s
<5% U
T
(>95% dips of
U
T
)
for 0,5 cycle
<5% U
T
(>95% dips of
U
T
)
for 1 cycle
70% U
T
(30% dips of
U
T
)
for 25 cycles
<5% U
T
(>95% dips of
U
T
)
for 5s
Mains power quality
should be at least that
one of a typical
commercial or hospital
environment.
If the user of the
MAG2000 requires
continued operating
during power mains
interruptions, it is
recommended that
MAG2000 be powered
from an uninterruptible
power supply (UPS) or a
battery
Mains power
electromagnetic
field (50/60Hz)
IEC 61000-4-8
30A/m
30A/m
Mains power quality
should be at least that
one of a typical
commercial or hospital
environment.
Note: U
T
is the AC mains voltage before the application of the Test level.