IACER Srl 26 MNPG69-09
ELECTROMAGNETIC COMPATIBILITY TABLES
Guidance and manufacturer's declaration ELECTROMAGNETIC EMISSIONS
FOR ALL DEVICES AND SYSTEMS
MAG2000 family is expected to operate in the electromagnetic environment
below specified. The customer or user of the MAG2000 family must ensure that
it is used in such environment.
Emission test
Compliance
Electromagnetic environment
guidance
RF emissions
CISPR 11
Group 1
MAG2000 family uses RF energy
only for its internal function.
Therefore, its RF emissions are
very low and are unlikely to cause
any interference in nearby
electronic equipment
RF emissions
CISPR 11
Class B
MAG2000 is suitable for use in all
establishments, including
domestic establishments and
those directly connected to the
public low-voltage power supply
network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Compliant
Guidance and manufacturer's declaration ELECTROMAGNETIC IMMUNITY
FOR ALL DEVICES AND SYSTEMS
MAG2000 is intended for use in the electromagnetic environment specified
below. The user or operator of MAG2000 should assure that is used in such
environment.
Immunity test
Test level
IEC 60601
Compliance
level
Electromagnetic
environment
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6kV; +8kV at
contact
±8kV; +15kV in
air
±6kV; ±8kV
at contact
±8kV; +15kV
in air
Floors should be wood,
concrete or ceramic
tile. If floor is covered
with synthetic material,
the relative humidity
should be at least 30%.
Electrical fast
transient/burst
±2kV
for power
supply lines
±2kV
for power
supply lines
Mains power quality
should be at least that
one of a typical