IACER Srl 38 MNPG148-06
Prg
Medical
prg
Yes/No
Description PHASE 1 PHASE 2 PHASE 3
Frequency 6
Hz
impulse
width 250µs
250µs x 8sec
CH1&CH2
80%+ 50Hz
250µs x 6 sec
75% CH1 +
50Hz 250µs x
6 sec 75%
CH2) x 45
cycles
Frequency 10
Hz
impulse width
250µs (80%)
15 No
Sequential
tonic
contractions
upper limbs
and trunk
Total time 3
min
Frequency 6
Hz
impulse
width 200µs
Total time 10
min (30Hz-
200
µs x 5 sec
80% CH1 +
30Hz-200µs x
5
sec 80%
CH2) x 60
cycles
Total time 5
min
Frequency 4
Hz
impulse width
200µs (90%)
16 No
Sequential
tonic
contractions
lower limbs
Total time 3
min
Frequency 6
Hz
impulse
width 300µs
Total time 10
min (30Hz-
300µs x 5 sec
80% CH1 +
30Hz-300µs x
5 sec 80%
CH2) x 60
cycles
Total time 5
min
Frequency 4
Hz
impulse width
300µs (90%)
17 No
Sequential
phasic
contractions
upper limbs
and trunk
Total time 3
min
Frequency 6
Hz
impulse
width 200µs
Total time 10
min (50Hz-
200
µs x 5 sec
75% CH1 +
50Hz-200µs x
5 sec 75%
CH2) x 60
cycles
Total time 5
min
Frequency 4
Hz
impulse width
200µs (90%)
18 No
Sequential
phasic
contractions
lower limbs
Total time 3
min
Frequency 6
Hz
impulse
width 300µs
Total time 10
min (50Hz-
300µs x 5 sec
75% CH1 +
50Hz-300µs x
5 sec 75%
CH2) x 60
cycles
Total time 5
min
Frequency 4
Hz
impulse width
300µs (90%)

Related product manuals