IACER Srl 46 MNPG148-06
MEM programs
Prg
Medical
prg.
Yes/No
Description PHASE 1
1-5 Yes
Free
memoriesTENS
Total time 1-90 min
frequency 1-200 Hz
width impulse 20-250 µs
6-10 No
Free memories
NEMS
Total time 1-90 min
frequency 1-200 Hz
contraction time 1-10 sec
slope 0-5 sec
recovery time 0-30 sec
width impulse 50-450
µ
s
11-12 No
Free memories
NEMS alternated
CH1/CH2
Total time 1-90 min
frequency 1-200 Hz
contraction time 1-10 sec
slope 0-5 sec
recovery time 0-30 sec
width impulse 50-450
µ
s
13
No
Battery test
M1-M5 • TENS Free memories (medical program)
Free memories for antalgic TENS treatment.
M6-M10 • NEMS Free memories (non-medical program)
Free memories for muscle recovery and training.
M11-M12 • NEMS Alternated free memories (non-medical program)
Free memories for muscle recovery and/or training with alternated impulses
on channel 1 and 2.
M13 • Battery test program (non-medical program)
Battery calibration program for the exclusive use of the manufacturer.