IACER Srl 57 MNPG148-06
Guidance and manufacturer’s declaration – ELECTROMAGNETIC IMMUNITY –
FOR ALL EQUIPMENT AND SYSTEMS
MIO CARE PRO is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in such
Immunity test
Compliance level
Electromagnetic
environment - guide
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
30A/m 30A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
T
is the A.C. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – ELECTROMAGNETIC IMMUNITY –
FOR EQUIPMENT AND SYSTEMS THAT ARE NOT LIFE-SUPPORTING
MIO CARE PRO is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in such
Immunity test
Electromagnetic environment
- guide
Portable and mobile RF communications equipment should not be used near any
part of the device (including cables) except when the
distance is respected. This distance is calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
eff
from 150kHz
to 80MHz
6V
eff
in ISM band
and radio
bands
between
150kHz and
eff
from 150kHz
to 80MHz
6V
eff
in ISM band
and radio
bands
between
150kHz and
d = 1,2
√
P
from 150kHz to 80MHz
d = 1,2
√
P
from 80MHz to 800MHz
from 80MHz
from 80MHz
Where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in metres (m).