IACER Srl 59 MNPG148-06
Recommended separation distances between portable and mobile RF
communications equipment for MIO CARE PRO that are not life-supporting
MIO CARE PRO is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of MIO CARE PRO
can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and
MIO CARE PRO as recommended below, according to the maximum output power
of the communications equipment.
Rated maximum
output power of
transmitter
(W)
Separation distance according to the frequency of the
transmitter (m)
from 150kHz
to 800 MHz
from 80MHz
to 800 MHz
from 800MHz
to 2,7GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be determined using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
Note
1) At 80 MHz and 800 MHz the separation distance for the higher frequency range applies.
2) These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.