IACER Srl 14 MNPG117-02
With ultrasound therapy inflammation temporary increases can happen in
treatment area, and so pain temporary increase, traumas due to more dosage,
nervous system reactions, sanguine coagulation. If this occurs, suspend the
treatment and consult a doctor.
Warnings
It is recommended:
to control position and meaning of all the labels and symbols on the
equipment;
the device doesn’t produce or receive electromagnetic interferences from
other devices. However, it’s recommended to keep a distance of at least 3
meters from televisions, monitor, mobile phones or other electronic
devices (which can lead to abnormal device behavior). Care must be taken
when operating this equipment around other equipment. Potential
electromagnetic or other interference could occur to this or to the other
equipment;
to avoid the use of the device by persons who did not read carefully this
manual;
not to wear metal objects during treatment (both for the operator and
both for the patient);
to use ONLY accessories supplied by device manufacturer. It is
recommended to use the device only with the supplied power supply
MPU50-160. I-TECH UT2 is tested and guaranteed for use with the
supplied accessories.
It is forbidden:
to use the device in the presence of MRI equipment and patient
monitoring equipment, of electrosurgical (possible bruises and burns) or
shortwave or microwave therapy equipment or other equipment that
sends electrical impulses into the body and in general in combination to
other medical devices;
to use the device by persons known to be unsound-minded, or suffering
from sensibility disorders, permanently or temporarily disabled unless
assisted by qualified personnel (e.g. a doctor or therapist); by persons
younger than 12 years old or not adequately educated about the device
use by an adult person;
ultrasounds should not be used on areas with reduced sensitivity or
circulation. Patients experiencing reduced sensitivity may not be able to
warn their therapist/doctor when the ultrasound is too intense. Patients
experiencing circulation problems may suffer from an excessive increase
of temperature in the treated area;