IACER Srl 32 MNPG117-02
ELECTROMAGNETIC COMPATIBILITY TABLES
Guidance and manufacturer’s declaration – ELECTROMAGNETIC EMISSIONS – FOR
ALL EQUIPMENT AND SYSTEMS
The I-TECH UT2 device is intended for use in the electromagnetic environment
specified below. The customer or the user of the I-TECH UT2 should assures that it is
used in such an environment.
Emission test Compliance
Electromagnetic environment -
Guidance
RF emissions
CISPR 11
Group 1
The I-TECH UT2 device uses RF energy
only for its internal function. Therefore,
its RF emissions are very low and are not
likely to cause any interference in nearby
Class B
The I-TECH UT2 device is suitable for use
in all establishments other than
domestic and those directly connected
to the public low-
network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
/ flicker emissions
Compliant
Guidance and manufacturer’s declaration – ELECTROMAGNETIC IMMUNITY – FOR
ALL EQUIPMENT AND SYSTEMS
The I-TECH UT2 is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in such an
Immunity test
Test level
IEC 60601
Compliance level
environment -
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8kV in contact
±15kV on air
±8kV in contact
±15kV on air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative
humidity should be
at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
for power
for power
Mains power quality
should be that of a
typical commercial
or hospital
environment.
for input/output
for input/output