Icare HOME - Setting Up the Tonometer; Probe Loading Steps

Icare HOME
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
www.icaretonometer.com
11
English
LOAD
MEASURE
REPEAT
DONE
SERVICE
BAT TERY
LOA
EASU
LO
EAS
1. Unwrap the probe.
2. Remove the lid of the
probe container as
shown in the gure
above. Point the
tonometer upward.
3. Drop the probe into
the probe base (1)
by turning the probe
container upside down.
4. Press the measurement
button (13) briefly (1
second) to activate the
probe.
5. The probe moves rapidly
back and forth.
6. See that the Measure
light (15) ashes. If so,
the probe is loaded
correctly and ready for
measurement.
To load the probe:
1X
Icare HOME tonometer Instruction manual for health care professionals
Icare HOME
Icare HOME tonometer Instruction manual for health care professionals ENGLISH
11
www.icare-usa.com