Guidance and manufacturer’s declaration IEC 60601-1-2: Edition 4.0; 2014-02 – Electromagnetic immunity
Icare HOME (TA022) is intended for use in a home healthcare environment with electromagnetic
characteristics specified below. The customer or the user of the Icare HOME (TA022) should
assure that it is used in such an environment.
Immunity test IEC 60601 test level
Compliance
level
Electromagnetic environment-Guidance
Proximity fields
from RF wireless
communications
equipment
IEC 61000-4-3
380 - 390 MHz 27 V/
m; PM 50%; 18 Hz
430 - 470 MHz 28 V/m;
(FM ±5 kHz, 1 kHz sine)
PM; 18 Hz
704 - 787 MHz 9 V/m;
PM 50%; 217 Hz
800 - 960 MHz 28 V/m;
PM 50%; 18 Hz
1700 - 1990 MHz 28
V/m; PM 50%; 217 Hz
2400 - 2570 MHz 28
V/m; PM 50%; 217 Hz
5100 - 5800 MHz 9 V/
m; PM 50%; 217 Hz
27 V/m
28 V/m
9 V/m
28 V/m
28 V/m
28 V/m
9 V/m
WARNING: Portable RF communications
equipment (including peripherals such as
antenna cables and external antennas) should
be used no closer than 30 cm (12 inches) to any
part of the
device1), including cables specified by the
manufacturer. Otherwise, degradation of the
performance of this equipment could result.
Interference may occur in the vicinity of
equipment marked with the following symbol: