Error messages
To clear error messages, press the measurement button, after which the measurement can be repeated.
The following messages may appear:
bAtt ........................... The batteries are low. Replace the batteries
E 01 ........................... The probe did not move at all.
If this error message is repeated, turn the tonometer so that the collar faces down for a short time.
If the error message is repeated, remove the probe base and replace it with new ones
(see page 9).
E 02 .............................The probe did not touch the eye.
T
he measurement was taken from too far away.
E 03 ..............................The probe speed was too low.
The measurement was taken from too far away or the tonometer was tilted too far upwards.
E 04 ..............................The probe speed was too high.
The tonometer was probably tilted downwards. Make sure that the central groove is in the
horizontal position.
E 05 ...............................The contact with the eye was too “soft.
The probe probably made contact with the eyelid.
E 06 ................................The contact with the eye was too “hard.
The probe made contact with the opening eyelid or calcification in the cornea.
E 07 ................................”Bad hit.
The probe measurement signal detected by the tonometer was unusual. The probe may have
made contact with a peripheral part of the cornea or the probe was twisted or otherwise inserted
incorrectly. If this error message repeats, remove and replace the probe.
E 09 ................................”Bad data.
An erroneous measurement for a reason other than those described in E01−E07.
Service procedures
Replace the batteries when the <bAtt> message appears.
Change the probe base if the probe does not move smoothly.
No other service procedures can be carried out by the user. All other servicing and
repairs must be carried out by the manufacturer or certified service facilities.
The device can be cleaned with a damp cloth containing disinfectant.