Maintenance
Follow local regulations and recycling instructions regarding the disposal or recycling of the Icare®
Tonometer and accessories.
Returning the Icare® Tonometer for servicing /repair
Contact Icare Finland’s Technical Services Department (see www.icarefinland.com) or your local Icare
representative for shipping instructions. Unless otherwise instructed by Icare Finland, there is no need
to ship accessories along with the tonometer. Use a suitable carton with the appropriate packaging
material to protect the device during shipment.
Return the device using any shipping method that includes proof of delivery.
Service
Diagram of tonometer functions
Tonometer ON
Ready for measurement
Old values search Turning OFF
Shutdown
OLD00 END 00
1.18
5.18
P.18
P0 18
P9 18
Earlier measurement
values (0-9)
Final result
Measurement button (measure, confirm)
Selector button (left and right)
WARNING
The tonometer should only be opened by qualified service personnel. It contains no user-
serviceable parts, apart from the batteries and a probe base.
The Icare® Tonometer requires no routine servicing or calibration other than changing the
batteries at least every 12 months or changing the probe base.
If servicing is necessary, contact qualified service personnel or your local Icare representative.
13
ALL manual_icare _A5_100209.indd 13ALL manual_icare _A5_100209.indd 13 2.3.2009 11:06:262.3.2009 11:06:26