Summary

Explanation of Warning Symbols

General Hazard Statement

General hazard statement indicating importance for health and safety.

Danger Symbol (Health/Safety)

Failure to observe this information may be detrimental to health or may result in injuries.

Warning Symbol (Technical)

Important for ensuring appliance functions without technical problems.

Caution Symbol (Operational)

Important for proper use and correct appliance function, leads to accurate results.

General Safety Instructions

Before First Use

Read operating instructions and follow safety guidelines before starting up.

Burn Risk Warning

Exercise caution when touching housing parts and heating plate due to high temperatures.

Flammable Media Warning

Caution: Only process media with a flash point higher than 180 °C.

Safe Temperature Limit Setting

Set safe temperature limit at least 25 °C lower than the media's fire point.

Hazardous Environments

Do not operate in explosive atmospheres, with hazardous substances, or under water.

Correct Use Guidelines

Commissioning Procedures

Safe Temperature Limit (RT)

Temperature Restriction

Max heating plate temperature restricted to 130 °C; device switches off heating above this limit.

Temperature Limit Warning

Set safe temperature limit at least 25 °C lower than the media's fire point.

Setting the Safe Temperature Limit

Procedure to Set Limit

Switch on, wait for SAFE display, press Temp key and use +/- keys to set.

Saving the Temperature Limit

The set safety temperature is displayed and saved upon releasing the Temp key.

Stirring Function Operation

Direction of Rotation Reversal

Heating Function Operation

Maintenance and Cleaning

Error Codes and Solutions (RT)

Error Er 3: High Temperature

Error when device temperature is too high. Solution: Switch off, cool down, check plug.

Error Er 6: Safety Circuit Break

Error due to break in safety circuit. Solution: Check connection, plug, or sensor.

Error Er21: Safety Relay Test Fail

Safety relay does not open during test. Solution: Switch off, check PCB (authorised personnel).

Error Er22: Low Safety Temperature

Safety temperature is too low during test. Solution: Switch off, check PCB/sensor (authorised personnel).

Error Er25: Heating/Switching Monitoring

Error in heating and switching element monitoring. Solution: Check controller, switch off, check PCB.

Error Er26: Temperature Difference

Difference between safety and control sensor exceeds limit. Solution: Switch off, check connector/PCB.

Error Er31: Triac Detection Error

Triac error detection is not working. Solution: Switch off, check PCB (authorised personnel).

Error Er46: Safety/Control Temp Diff

Difference between safety and control sensor exceeds limit. Solution: Switch off, check connector/PCB.

Troubleshooting Beyond Codes

If faults persist or other codes appear, contact service or send for repair.