WARRANTY AND REPAIR
5| Either arc amount or arc time does not change.
•
On SM, NZ, MM or AUTO mode, either arc amount or arc time does not change. Implement [Arc
Calibration] and the arc amount on these modes properly maintain. When used in another mode, arc
amount and arc time are automatically set to prevent their alteration.
6| Set pause.
•
Refer to [Option menu].
7| Indicate cleaved angle, fiber angle, and clad deviation.
•
Refer to [Option menu].
8| Estimated splice loss and measured splice loss do not match.
•
The estimated splice loss is a calculated value so it should be used only as a reference.
9.3
| SLEEVE HEATER
1| Fiber protecting sleeve does not contract completely.
•
Increase the heating time. Refer to [Heater mode edition].
2| The heater is overheated.
•
Stop the heater by pressing , turn the power off and then contact UCLSWIFT.
•
If the protecting sleeve melts and sticks to the heater cover, remove it by pushing it with a cotton swab.
3| Initialize heater mode condition.
•
Refer to [Heat mode edition].
4| Cancel heater in the middle of an operation.
•
Heater operation cannot cancel by pressing . Cancel it by pressing once again.
9.4
| OTHERS
1| Restrict splice mode and heater mode setting.
•
Refer to [Menu lock].
2| Splice mode’s arc amount does not change even after [Arc calibration].
•
The internal standard arc amount does calibrate. Therefore, the arc amount of each splice mode does
not change.
Responsibility limit
For Warranty and Repair information, please contact your UCLSWIFT representative or use the contact
information found in the Help section under the information menu in the splicer unit.
10.1
| INFORMATION NECESSARY FOR REPAIR
Before sending the product, contact UCLSWIFT first.
1| Company and contact information (Name, department, company, address, contact information, fax, e-mail)
2| Product serial number
3| Product condition and problem incurred, error information
4| Operating conditions, processes and uses of the splicer prior to error
10.2
| TRANSPORTATION
Please return the splicer in the carrying case provided to protect it from humidity, vibration and shock. Include
all components in the case.
10.3
| REPAIR
Customized modes, configurations and splice data may be lost during repair. Save test data prior to shipment
to ensure retention.
3| Forgotten password.