OPERATING IN REMOTE CONTROL
CHPT 3
PAGE 10
7 6 5 4 3 2 1 0
7 6 5 4 3 2 1 0
15
14
13 12
11 10
8
9101112131415
9
8
Logical
OR
Turn Output Off
LASER Output Off
Register
LASER Output Off
Enable Register
LASer:ENABle:OUTOFF <nrf>
LASer:ENABle:OUTOFF?
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
Bit Reference
0- LASER Current Limit 8- N/A
1- LASER Voltage Limit 9- LASER Output Changed to be out of Tolerance
2- N/A 10- N/A
3- LASER Power Limit (output on) 11- N/A
4- N/A 12- N/A
5- N/A 13- N/A
6- N/A 14- N/A
7- N/A 15- N/A
Figure 3.3 LDX-3200 Series' Output Off Register
Command Timing and Completion
This section describes, for each device-dependent command, whether that command is performed in an overlapped or
sequential manner. In other words, it states whether the next command may begin while this command is being
executed, or if the next command must wait until this command is completed before its execution begins. The
conditions for setting the operation complete flag are given in the Chapter 3 section titled Operation Complete
Definition.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com