OPERATING IN REMOTE CONTROL
CHPT 3
PAGE 12
LDX-3200 Series ERROR MESSAGES
Error Code Explanation
E-001 Memory allocation failure.
E-103 <DEFINITE LENGTH ARBITRARY BLOCK PROGRAM DATA> length too long.
E-104 <NON-DECIMAL NUMERIC PROGRAM DATA> type not defined.
E-105 <DECIMAL PROGRAM DATA> exponent not valid.
E-106 <DECIMAL PROGRAM DATA> digit expected.
E-123
E-124
Command not found.
<program mnemonic> Lookup, failed because query/command type match failed.
E-126 Too few or too many program data elements.
E-201 <PROGRAM DATA> value out of range.
E-202 <PROGRAM DATA> will not convert to valid type.
E-203 Security violation, command is not available without clearance.
E-205 <PROGRAM DATA> is not a Boolean value or word.
E-207
E-209
<PROGRAM DATA> will not convert to an unsigned 16-bit value.
<PROGRAM DATA> will not convert to an unsigned 32-bit value.
E-210 <PROGRAM DATA> will not convert to a floating point value.
E-211 <PROGRAM DATA> will not convert to a character value.
E-213 <PROGRAM DATA> is incorrect block data length.
E-214
E-302
<PROGRAM DATA> length exceeds maximum.
Query error. Device was addressed to talk, but GPIB Controller failed to read all of the <RESPONSE
MESSAGE>.
E-501 LASER interlock disabled output.
E-503 LASER open circuit disabled output.
E-504
E-505
LASER current limit disabled output.
LASER voltage limit disabled output.
E-507 LASER Power limit disabled output.
E-510 LASER out of tolerance disabled output.
E-511 LASER control error disabled output.
E-512 Analog section status is all 1's or all 0's (power down).
E-513 Serial EEPROM checksum error.
E-515 Laser Output must be off to change ranges.
E-516 Incorrect Configuration for Calibration Sequence to start.
E-519 Setting a measurement is only valid during the calibration phase for that measurement. User has tried to
calibrate a measurement without first entering the required calibration mode.
Table 3.5 LDX-3200 Series Error Codes