Questions and Answers

  • K
    Kristen MeyerAug 18, 2025
    What to do if the Voltage Limit indicator blinks on my ILX Lightwave Portable Generator?
    • T
      Thomas JamesAug 18, 2025
      If the ILX Lightwave Portable Generator voltage limit indicator is blinking, check the laser connections. This condition may occur due to high impedance, especially at high current levels.
  • C
    Christopher MartinAug 20, 2025
    What to do if my ILX Lightwave LDX-36000 Portable Generator shows an Open Circuit error E503 or Voltage Limit error E505?
    • C
      Crystal SanchezAug 20, 2025
      If your ILX Lightwave Portable Generator displays an Open Circuit error (E503) or Voltage Limit error (E505) that prevents the output from reaching the desired value, check if the voltage limit setting is too low, as the LDX-36000 series has an adjustable compliance voltage. Also, inspect the laser connections.
  • T
    Tony VazquezAug 22, 2025
    What to do if my ILX Lightwave LDX-36000 Portable Generator display is frozen?
    • J
      Justin PhillipsAug 23, 2025
      If the ILX Lightwave Portable Generator is powered on, but the display is frozen and the buttons do not operate, turn the power switch off, and then back on to restart it. This issue might happen if the instrument loses power briefly. Also, make sure the instrument is not in Remote mode by pressing the Local button, and then try changing the display mode.
  • P
    Paul BrooksAug 25, 2025
    How to troubleshoot power issues with my ILX Lightwave Portable Generator?
    • N
      Nathan HartAug 26, 2025
      If the ILX Lightwave Portable Generator is not powering up, check the AC power line voltage, the power cord connection, and the fuse.
  • D
    Daniel TaylorAug 28, 2025
    Why does the output of my ILX Lightwave Portable Generator go off intermittently?
    • J
      Jesse PerezAug 28, 2025
      If the ILX Lightwave Portable Generator output goes off intermittently, an intermittent interlock may be disabling the output. Verify the interlock connections and ensure the AC power cord connection is secure.
  • E
    Ebony ShahAug 30, 2025
    What to do if my ILX Lightwave Portable Generator has no output in optical power mode?
    • C
      Christina BryantAug 30, 2025
      If there is no output in optical power mode on your ILX Lightwave Portable Generator, even though the front panel indicates the laser should be outputting light, and the laser is outputting light even though the front panel says it should be below threshold, verify the correct values for threshold and slope efficiency have been entered. Check by observing the values stored for the parameter Cal P.
  • K
    kelly78Sep 2, 2025
    What to do if my ILX Lightwave LDX-36000 Portable Generator has power but no output?
    • C
      Christina SweeneySep 2, 2025
      If the ILX Lightwave Portable Generator is powered on but there is no output, verify that the interlocks are satisfied, the output cable is securely connected to the output connector, and the connections to the laser are secure.
  • U
    uthompsonSep 4, 2025
    Why does the measured voltage always read zero on my ILX Lightwave LDX-36000 Portable Generator?
    • J
      Julie SmithSep 4, 2025
      If the measured voltage always reads zero on your ILX Lightwave Portable Generator, verify that the voltage sense lines are connected across the load: Vsense+ to laser anode and Vsense- to laser cathode.
  • L
    lawrencedominiqueSep 7, 2025
    Why is the temperature display incorrect on my ILX Lightwave LDX-36000?
    • H
      Holly SandersSep 7, 2025
      If the temperature cannot be displayed or the measurement is incorrect on your ILX Lightwave Portable Generator, verify that the Steinhart-Hart constants have been set to non-zero values and are appropriate for the thermistor being used.
  • T
    thompsonjacquelineSep 9, 2025
    How to set or control the temperature on my ILX Lightwave LDX-36000?
    • S
      Shannon LittleSep 9, 2025
      Temperature is a measurement only and cannot be controlled from the ILX Lightwave LDX-36000 Portable Generator.