CALIBRATION AND TROUBLESHOOTING
Error Messages
136 LDX-36000 Series
CHAPTER 6
Table 6.3 Error Messages
Error Code Error Message
E-001 Memory allocation failure.
E-103 <DEFINITE LENGTH ARBITRARY BLOCK DIAGRAM DATA> length too long.
E-104 <NON-DECIMAL NUMERIC PROGRAM DATA> type not defined.
E-105 <DECIMAL PROGRAM DATA> exponent not valid.
E-106 <DECIMAL PROGRAM DATA> digit expected.
E-124 <PROGRAM MNEMONIC> Lookup, failed because query/command type match failed.
E-126 Too few or too many program data elements.
E-201 <PROGRAM DATA> value out of range.
E-202 <PROGRAM DATA> will not convert to valid type.
E-203 Security violation, command is not available without clearance.
E-205 <PROGRAM DATA> is not a Boolean value or word.
E-207 <PROGRAM DATA> will not convert to an unsigned 16-bit value.
E-210 <PROGRAM DATA> will not convert to a floating point value.
E-211 <PROGRAM DATA> will not convert to a character value.
E-213 <PROGRAM DATA> is incorrect block data length.
E-214 <PROGRAM DATA> length exceeds maximum.
E-301 A <RESPONSE MESSAGE> was ready, but controller failed to read it. (query error)
E-302 Query error. Device was addressed to talk, but controller failed to read all of the
<RESPONSE MESSAGE>.
E-501 Laser interlock #1 disabled output.
E-502 Laser interlock #2 disabled output.
E-503 High impedance condition.
E-504 Laser current limit disabled output.
E-505 Laser voltage limit disabled output.
E-506 AC power failure
E-509 Laser output disabled because of High Temperature Limit condition.
E-511 Laser control (unknown) error disabled output.
E-525 Open temperature sensor error.
E-526 Shorted temperature sensor error.
E-527 Power supply failure.
E-528 Power supply voltage limit.
E550 Pass element power limit
E-599 Laser open circuit 2.