CALIBRATION AND TROUBLESHOOTING
Calibration Overview
03_10 LDX-36000 Series 125
CHAPTER 6
Beginning Calibration
To begin the calibration process, turn on the 36000 and allow it a warm-up period
of 60 minutes prior to performing any calibration.
Once this warm-up period is over, connect the 36000 to a GPIB compatible
computer.
The default GPIB address for the 36000 is 1. This can be changed on the front
panel of the 36000. All calibrations require slope and offset values to be sent to
the 36000.
Be sure to reset the calibration to default values (1,0) before each calibration is
performed.