Electromagnetic Immunity
The device is intended for use in the electromagnetic environment specified below. The user
must assure that the device is used in such an environment.
Electromagnetic
Environment - Guidance
Electrostatic
discharge
(ESD)
IEC 61000-
4-2.
± 8KV for contact;
± 2KV, ±4 KV, ±8
KV, ± 15KV for air
± 8KV for contact;
± 2KV, ± 4KV, ± 8KV,
±15 KV for air
Floors must be wood,
concrete, cement or tiled.
If floors are covered with
synthetic material (carpet,
etc.), the relative humidity
must be at least 30%.
Electrical
fast
transients
IEC 61000-
4-4.
±2KV for power
supply lines; ±1 KV
for interconnect lines
±2KV for power supply
lines; ±1 KV for
interconnect lines
Mains power quality
should be that of a typical
commercial or hospital
environment (hospital,
clinic).
±0.5KV, ±1KV (line
to line); ±0.5KV,
±1KV, ±2KV (line to
earth)
±0.5KV, ±1KV (line to
line); ±0.5KV, ±1KV,
±2KV (line to earth)
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage
dips, short
interruptions
and voltage
variations
IEC 61000-
4-11.
0% residual voltage
for 0.5 cycle.
0% residual voltage
for 1 cycle.
70% residual voltage
for 25 cycles.
0% residual voltage
for 250 cycles.
0% residual voltage
for 0.5 cycle.
0% residual voltage
for 1 cycle.
70% residual voltage
for 25 cycles.
0% residual voltage
for 250 cycles.
Mains power quality
should be that of a typical
commercial or hospital
environment.
If the use of the device
requires continued
operation during power
mains interruption, it is
recommended that the
product be powered from
a separate power supply
(UPS, etc.).