EasyManua.ls Logo

iM3 SP6 - Section IV. Infection Control

iM3 SP6
118 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
- 9 -
SECTION IV
INFECTION CONTROL
Infection Control:
Always use standard personal protective equipment such as facemasks, eyewear, face shields,
gloves, and protective gowns.
In addition to using protective equipment, ensure that your water supply system conforms to
applicable Centers for Disease Control and Prevention (CDC) and American Dental Association
(ADA) standards, and that all required flushing, chemical flushing, and general infection control
procedures are followed.
To further protect against infection, using distilled or purified water is recommended.
Cleaning and Sterilization:
All sterilization procedures must be followed in accordance to the EN ISO17664 standards.
In order to avoid bacterial or viral infections, always clean and sterilize the following components
after each treatment:
1. Hand piece
2. Tips
3. Metal Wrenches
4. Micro-motor
5. Straight hand piece (Straight nose cone)
6. Prophy head
The hand piece and tips, wrench, straight handpiece and prophy head can be autoclaved and can
withstand a maximum temperature of 135 or (275) for a maximum of 3 minutes. After
completing the sterilization cycle, air-dry for about 20-30 minutes before reusing.
WARNING: Do not use any other method of sterilization (dry heat, radiation, ethylene oxide, gas,
low-temperature plasma, etc.)
NOTE
The Sterilization Assurance Level (SAL) of steam autoclave should be 10
-6
(According
to ISO 13683:Sterilization of health Care products)
NOTE: Prior to cleaning and sterilizing your ART-SP6 unit, always remember to turn off the
device by using the power switch and to disconnect the power plug from the outlet.

Related product manuals