- 26 -
SECTION X: SPECIFICATIONS
The ART-SP6 scaler is designed and manufactured to meet the most demanding environment. It’s
specifications are listed below
Classification
Protection against electric shockClass I
Degree of protection against electric shockType BF
Mode of operationContinuous
Medical device directiveIIa
Standards Followed
ISO13485: Medical Devices, Quality Management Systems, Requirements for regulatory
purposes
93/42/EEC: MDD CE Directive
EN60601-1: Medical Electrical Equipment, General Requirement for Safety
EN60601-1-2: Medical Electrical Equipment, Electromagnetic Compatibility,
Requirements and Tests
Specification
Power supply
115V ±5%
50/60Hz 125VA
230V ±5% 50/60Hz 125VA
Working frequency
26KHz ~ 32KHz
Polisher speed
3000
30000 R.P.M
Water supply
25
60 PSIG(172~414KPa)
Dimension
23cm(L) x 21cm(W) x 9cm(H)
Weight
3.7 Kg (including the handpiece)
Hand piece Cable
250 cm
Footswitch Cable
250 cm
Operation environment
Temperature
0 ~40
Relative Humidity
10% ~ 90% (non-condensing)
Transport and storage conditions
Temperature
0
℃~ 60℃
Relative Humidity
10% ~ 90% (non-condensing)
Atmospheric pressure
860~1060 hPa
SECTION XI: DISPOSAL
Please follow county and state regulations for disposal of the ART-SP6.