4
4)
Miscellaneous
a) No Ready Light: Make sure that the jumper between 8 & 9, on the bar furthest from the Power Switch, is in
place.
b) Machine Doesn’t Work but Everything Else is Fine: Check that J1 is to the right and covers pins 1 & 2. J1
location determines whether the Hand-Held device is used or whether an external device is used (J1 is located
on the opposite side to the Power Switch).
c) Setting the System in Remote Control Mode: Move J1 to the right so it covers the left and middle pins (orient
the control box so you would be able to read the label next to the Power Switch.
d) Machine Just Clicks but doesn’t operate: Check that K2 on the PC board is in place (located on the upper-right
corner when you orient the control box so you are able to read the label next to the Power Switch). Check that
the connections on the hand-held control are in place as well.
e) For duel tube head operations place a jumper between 14 & 17 on bar M2.1 (The one closest to the Power
Switch). In addition, connect:
Number 13 on bar M2.1 to tube head B and to wall mount.
Number 11 on bar M2.1 to tube head B and to wall mount.
Number 10 on bar M2.1 to tube head A and to wall mount.
The micro-switches provided are to be connected as follows:
-Tube A is connected between terminals 8 & 9 on the M1 bar in the control box and the lower terminal 8 & 9 on
the steel mounting plate.
-Tube B is connected between terminals 8 & 10 on the M1 bar in the control box and the lower terminal 8 & 9
on the steel mounting plate.