TROUBLE SHOOTING TX
TX
hexadecimal Error description details solution
TX
01 STOP button failure
Microswitch contacts of STOP button do not
open synchronously
In case of repeated error, replace the
STOP button
11 Battery voltage out of range for the logic board
The voltage is higher than 3,2V (risk of board
damage)
Replace the battery with a genuine one
or correctly charged
Batter
totall
em
t
remainin
char
e
The voltage is lower than the minimum
Re
lace the batter
with a char
ed one
insufficient to start properly the TX)
(Wave=2Vdc Thor/Zeus=3Vdc)
20
a) Active command at Start (only for Wave)
b) bad comunication between micro’s
a) Not allowed at Start command
b) Time out in micro’s data exchange
a) Deactivate the command
b) Replace TX logic board
28 Bad comunication between micro and eeprom Probable damage along the comunication path Change the mother board
START push button is mechanically blocked
30 START command blocked
(dirt or broken spring)
Clean or change the STARTcontact
C5 Joystick failed
“forward direction” and “analogue” sensing do
not match; or the joystick is unplugged; or the
o
stick has been dama
ed inside
Check the joystick plug; replace the
joystick
11
C6 Joystick failed
“backward direction” and “analogue” sensing
do not match; or the joystick is unplugged; or
the joystick has been damaged inside
Check the joystick plug; replace the
joystick