7
•
If this equipment is modified, appropriate inspection and testing must be conducted to ensure continued safe
use of equipment.
•
Do not directly contact the InBody270 with any other electronic device when the InBody270 is on.
This may result in an electric shock.
DANGER
NOTECAUTION
Caution
If the InBody270 is not plugged into a grounded outlet, it may cause damage through electric surges or
product malfunction. This may affect the test results.
The test results may be inaccurate if the InBody270 is under electrical interference. Do not install the
InBody270 near products that generate electrical interference such as fluorescent lights, large AC motor
equipment(treadmill, vibration plate, refrigerator, air-conditioner, compressor, etc.), high-frequency thermal
therapy equipments, or heating appliances. Do not share the power source of the InBody270 with other
electrical devices. This may affect the test results.
When connecting the InBody270 with other test equipment, turn on the other equipment first. When
turning off other equipment, turn off the InBody270 first. This is necessary to minimize electrical surges
on the InBody270.
Always use the specified adapter provided by InBody as it is a part of the InBody270. Using other adapters
may result in malfunction of the InBody270.
Operation of the InBody270 2,000m above sea level may affect the weight measurement.
Adapter must be arranged so that easily cut off the power when a problem occurs in the InBody270.
5. Flip the power switch to turn on the InBody270.
InBody
Power switch
270_manual_ENG_B_150608.indd 7 2015. 6. 8. 18:45