Please test Element before fitting the patient with the prosthesis. Connect all
parts together and power ON the prosthesis. Connect to the software and tap
on each electrode. You should observe peaks on the signals and the terminal
device should move.
Testing Element
5
Note: The patient’s first use of the device must be in the presence of
a trained practitioner.
MAINTENANCE
Cleaning Electrodes
For best signal responses from your EMG electrodes, we recommended cleaning
the electrodes regularly to remove sweat residue and maintain hygiene. Clean
the surface of each electrode and the inner socket with a mild soap solution as
needed, at least a couple of times a week.
WARNING: DO NOT SUBMERGE whole electrode or arm in a bath of
water. This could permanently damage the electrodes.
WARNING: DO NOT USE harsh chemicals, such as acetone, bleach,
kitchen cleaners, etc. This could permanently damage the electrodes
and the skin.
Maintaining
Element
6
IBT Electrode and FlexCell Battery Replacements
IBT Electrodes and FlexCell batteries can be easily replaced if they are damaged
or stop working. If you need to replace the battery or electrodes, simply
unplug the battery or electrodes from the Element input connectors. Follow
the instructions in Section 4 to plug in the new electrode or battery. For any
abnormal issues, discontinue use and contact IBT for support.
The Element system components should not be thrown away with
common household waste. Dispose of Element components by either
returning the unit to IBT or taking the unit to an official electronics
disposal site.
DISPOSAL
REPAIRS, RETURNS, AND WARRANTY
Please contact IBT at support@i-biomed.com regarding repairs and returns.
The Element signal processing box and IBT Electrodes come with a 1-year
manufacturer’s defect warranty. Details of the warranty are enclosed separately.
TROUBLESHOOTING
For more detailed information, refer to the Element Troubleshooting guide.