Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The Element System is intended for use in the electromagnetic environment specified below.
The customer or the user of the Element System should assure that it is used in such an
environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF Emissions
CISPR 11
Group 2
The Element System must emit electromagnetic
energy in order to perform its intended function.
Nearby electronic equipment may be affected.
RF Emissions
CISPR 11
Class A
The Element System is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Not applicable
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Element System is intended for use in the electromagnetic environment specified below.
The customer or the user of the Element System should assure that it is used in such an
environment.
Immunity Test
IEC 60601 Test
Level
Compliance
level
Electromagnetic Environment -
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
±6 kV
contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
Not
applicable
Not applicable
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Not
applicable
Not applicable
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 s
Not
applicable
Not applicable
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 30 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a
typical commercial or hospital
environment.
NOTE: U
T
is the A.C. mains voltage prior to application of the test level.