Recommended separation distances between portable and mobile RF
communications equipment and MoMe
®
Kardia System
The MoMe
®
Kardia is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The user of the MoMe
®
Kardia can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications
equipment (transmitters) and the MoMe
®
Kardia as recommended below,
according to the maximum output power of the communications equipment
Rated maximum
output power
of transmitter
W
Separation distance according to frequency of transmit
150 kHz
to 80 MHz
d = 1.2 √ P
80 MHz
to 800 MHz
d = 1.2 √ P
800 MHz
to 2,5 GHz
d = 2.3 √ P
0.01
0.12
0.12
0.23
0.1 0.38
0.38
0.73
1
1.2
1.2
2.3
10 3.8
3.8
7.3
100 12
12
23
SAR Exposure Information
This device is designed and manufactured not to exceed the emission limits for exposure
to radio frequency (RF) energy set by the Federal Communications Commission of the
U.S. Government.
The exposure standard employs a unit of measurement known as the Specific Absorption
Rate, or SAR. The SAR limit relevant for the application described in the manual is 1.6W/kg.
Tests for SAR are conducted using standard operating positions accepted by the FCC with
the device transmitting at its highest certified power level in all tested frequency bands.
Although the SAR is determined at the highest certified power level, the actual SAR level
of the equipment while operating can be well below the maximum value. This is because
the device is designed to operate at multiple power levels so as to use only the power
required to reach the network. In general, the closer you are to a wireless base station
antenna, the lower the power output.
Equipment Authorization has been granted to this device with the reported SAR level(s)
evaluated as in compliance with the FCC RF exposure guidelines. SAR information on this
equipment is on file with the FCC and can be found under the Display Grant section of
www.fcc.gov/oet/ea/fccid after searching on the FCC ID as printed on the equipment.
This device has been tested to comply with FCC radiation exposure limits set forth for
an uncontrolled environment when used for the documented intended purpose and when
operated as shown in the user instructions provided with this product, i.e. when carried
with the belt clip coming with the product as a bundled accessory.
The MoMe
®
Kardia Device must be used with the included Belt Clip to keep a safe
minimum distance towards your body (6mm), ensuring compliance with regulatory
limits regarding human exposure to radio frequency radiation.