96-06434-00-01 revA
Service Manual, Inogen One G3HF
-85-
Figure 5: Detach compressor cable from motherboard
2. Carefully disconnect the feed waste manifold harnesses and the product manifold
harness from the motherboard using pliers to pull the white connectors off of the pins
and away from the motherboard (Figures 6 and 7).
Figure 6: Feed Waste Manifold Harnesses Figure 7: Product Manifold Harness
3. Remove the Accumulator Sensor Tubing by pushing down on the retaining ring while
gently pulling on the tubing. Be careful not to apply any force to the pressure sensor.
(Figure 8)