96-06917-00-01 rev C DCR 20-328 INOGEN ONE G4 TECHNICAL MANUAL
Page 14 of 19
4. INOGEN ONE G4 SYSTEM SPECIFICATIONS
Dimensions:
5.91”L (15.01cm) / 2.68”W (6.8 cm) / 6.45”H (16.3 cm)
5.91”L (15.01 cm) / 2.68”W (6.8 cm) / 7.15”H (18.2 cm) with
single battery
Weight:
2.8 lbs (1.27 kg.) (includes single battery)
Noise:
40 dB on setting 2
Warm-Up Time:
2 minutes
Oxygen Concentration:
90 +6/-3%
Flow Control Settings:
3 settings; 1-3
Power:
AC Power Supply
Input: 100-240 V, 50-60 Hz, 1.5 A
Output: 19 V, 3.42 A
DC Power Cable
Input: 13.5-15.5 V, 10 A
Rechargeable Battery: 12.0-16.8 V
Environmental Ranges
Intended for Use:
Temperature: 41 - 104
o
F (5 - 40
o
C)
Humidity: 0% to 95%, non-condensing
Altitude: 0 - 10,000 ft. (0 - 3048 m)
Environmental Ranges
Intended for Storage:
Temperature: -13 - 158
o
F (-25 - 70
o
C)
Humidity: 0% to 95%, non-condensing
Altitude: 0 - 10,000 ft. (0 - 3048 m)
4.1. Classifications
Mode of Operation:
Continuous
Type of Protection Against Electrical Shock:
Class II
Degree of Protection Against Electrical Shock:
Type BF
Not intended for cardiac application
Degree of Protection Against Ingress of Water:
IP 22
Degree of Safety for Application in Presence of
Anesthetic Gases:
Not suitable for such applications

Other manuals for Inogen One G4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Inogen One G4 and is the answer not in the manual?

Inogen One G4 Specifications

General IconGeneral
Weight2.8 lbs (1.27 kg) with single battery
Battery Life (single battery)Up to 2.7 hours
Battery Life (double battery)Up to 5 hours
TypePortable oxygen concentrator
Altitude RangeUp to 10, 000 feet (3, 048 meters)
Flow Settings1 to 3
PowerAC power supply 100-240V, 50-60Hz; DC power cable for mobile use
Oxygen FlowPulse Dose Delivery System
Oxygen Purity90% (+6% / -3%)
Operating Temperature41°F to 104°F (5°C to 40°C)
Storage Temperature-4 to 140°F (-20 to 60°C)