Inogen One® G4 System Specifications Chapter 8
Inogen One® G4 Concentrator
Dimensions:
With single battery
L / W / H : 5.91 in. (15.01 cm.) / 2.68 in. (6.8 cm.) / 6.5 in. (16.3 cm.)
L / W / H : 5.91 in. (15.01 cm.) / 2.68 in. (6.8 cm.) / 7.2 in. (18.2 cm.)
Weight: 2.8 pounds (1.27 Kg.) (includes single battery)
Noise: 40 dBA (on setting 2)
Warm-Up Time: 2 minutes
Oxygen Concentration: 90% - 3% /+ 6% at all settings
Flow Control Settings: 3 settings: 1 to 3
Power:
AC Power Supply:
DC Power Cable:
Rechargeable Battery:
AC Input: 100 to 240 VAC
50 to 60 Hz
Auto-Sensing: 1.0A
DC Input: 13.5-15.5VDC,10A Max.
Voltage: 12.0 to 16.8 VDC
Battery Duration*: Up to 2.7 hours with single battery
Up to 5 hours with double battery
Battery Charging Time: Up to 3 hours for a single battery
Up to 5 hours for a double battery
Environmental Ranges
Intended for Use:
Temperature: 41 to 104˚F (5 to 40˚C)
Humidity: 0% to 95%, non-condensing
Altitude: 0 to 10,000 ft (0 to 3048 meters)
Environmental Ranges
Intended for Shipping and
Storage:
Temperature: -13 to 158˚F (-25 to 70˚C)
Humidity: 0% to 95%, non-condensing
Store in a dry environment
Transportation: Keep Dry, Handle With Care
Tested by Independent
Laboratory:
Safety: IEC 60601-1
CAN/CSA C22.2 No. 60601-1
Electromagnetic Compatibility: IEC 60601-1-2, RTCA DO 160
Inogen One® G4
System Specifications
8
*Battery time varies with flow setting and environmental conditions.