E
edward22Jul 27, 2025
What to do if Inogen Medical Equipment concentrator does not power on?
- EEdwin FarrellJul 27, 2025
If your Inogen medical equipment concentrator doesn't power on, there are several potential causes. First, the battery might be discharged or not present, so use an external power supply or replace it with a fully charged battery. Second, the AC power supply might not be connected properly; check the connection and verify the green light is solid. Third, the DC cable may not be correctly connected; check both ends. If none of these steps work, it may be a malfunction, and you should contact your equipment provider.