Chapter: Requirements and Specifications
2-8 M10-16250-EN
Because of the wide range of applications where our instruments are used, Instron has no
knowledge of the environmental impact of the customer's test pieces or specimens.
Noise Levels
Noise levels do not include noise from specimen breaks.
Power Requirements
Table 2-2. Series 5980 Noise Levels
Maximum Noise Level dB(A)
Test Speed 5982 5984 5985 5987 5989
Standby or low speed 55 55 55 58 58
Maximum rated speed 72 74 72 72 72
Table 2-3. Series 5980 System Power Requirements
Parameter
Specifications
5982
a
5984
a
5985
a
5988 5989
Maximum Power
(VA)
3500 4800
Voltage (Vac)
(±10%)
200/208, 220, 230, 240
Single Phase
208/120, 240/139, 400/230, 480/277
3 Phase
Recommended
Circuit Breaker
b
20 Amp Type C 3-pole Type C circuit breaker, each rated at:
20 Amp Type C for voltage 208/120
or
16 Amp Type C for all other voltages
Frequency 47 to 63 Hz
a. Although these frames can operate at the indicated voltages, high speed turn-around times will be affected if voltage is
near extreme of tolerance.
b. A Type C circuit breaker is required between the facility power source and the load frame to accommodate high in-rush
currents at switch on.