The Generator is intended for use in the electromagnetic environment specified below. The customer or user of the Generator should
ensure that it is used in such an environment.
Maximum Output
Power (Was)
Separation (m)
150 kHz to 80 MHz
d = 1.2√P
Separation (m)
80 to 800 MHz
d = 1.2√P
Separation (m)
800 MHz to 2.5 GHz
d = 2.3√P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
Table 5—Recommended Separation Distances for the Generator, per IEC 60601-1-2, 3rd edition
TECHNICAL SPECIFICATIONS AND ENVIRONMENTAL REQUIREMENTS
WARNING: Per IEC 60601-1-2, 4th edition, portable RF communications should be used no closer than 30 cm (12 inches) to any part of the
Generator, including cords specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.