CUSA EXcel
Console Testing to IT0032 Issue 2
5.2 Bias 24v Duel DC power supply
24Volt Duel DC power supply output
(1)
Measure at J13
Set to 24Volts
Tolerances 23.5v to 24.5v
‘C’ Type Ultrasonic PCB
There are no adjustments to be
made on these power supplies