CUSA EXcel
Console Testing to IT0032 Issue 2
5.3 Vacuum Measurements
Digital
Analogue
With aspiration set to 100% and the
console in the RUN state
Connect tubing between the vacuum
port of the contamination guard and the
gauge
Vacuum level must be > 13in/Hg
Typical readings are 23 to 26in/Hg
Console Full Function Testing