CUSA EXcel
Console Testing to IT0032 Issue 2
5.4 Irrigation Measurements
Set 3cc/min on the irrigation display in the
STANDBY state
Connect an irrigation tube from an
irrigation source in to a measuring jug
Switch the console to the RUN state and
with a stop watch or timer run for 60
seconds then return the console to the
STANDBY state
Irrigation flow rate should be
3cc/min +/- 20%
Empty the contents of the measuring jug