CUSA EXcel
Console Testing to IT0032 Issue 2
5.5 Cooling flow rate Measurements
With the console in the STANDBY or RUN
state read the value on the cooling water
flow rate jig
Flow rate value must be > 35cc/min
Cooling flow rates are adjusted when setting
the cooling pump motor speed (5.1)
Console Full Function Testing