AT235 Additional Information Page 8
2.4 Reflex Latency
The reflex latency test will be automatically performed with an activator intensity of 10dB above the reflex
threshold. The test will show a pop up and ask for the intensity of the activator in case:
the reflex threshold cannot be found within the same protocol
the required intensity is at or above the warning level as determined in the protocol settings
the required intensity is above the maximum intensity that the transducer will permit for this
particular activator
The default display of the reflex latency test shows the graphs of the latency measurements from the
selected ear. The following information is found in the display:
A
The first 300ms of the tympanometric curve.
B
The table with measurement values which are only calculated if the measurement can be
completed.
Level, stimulus level
Pressure, the pressure at which the decay reflex is measured. Usually the decay test will
be set up to use the peak pressure of a preceding tympanogram.
Stimulus, stimulus frequency
Latency Value, the time interval between onset of the stimulus and the point where 10% of
the reflex deflection value is reached. The reflex deflection value is measured as the
average of the deflection between 250 and 300ms after stimulus onset.

Other manuals for Interacoustics AT235

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Interacoustics AT235 and is the answer not in the manual?

Interacoustics AT235 Specifications

General IconGeneral
BrandInteracoustics
ModelAT235
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals