TRV chair Instructions for Use - EN Page 33
5 Maintenance and care
The performance and reliability of the Vertigo Treatment and Rehabilitation (TRV) chair will be prolonged
if the following recommendations for care and maintenance are adhered to:
Regular:
Inspect the condition of the leg strap and the headrest and change the Velcro closures at
the first signs of wear
Inspect the condition of the foam on the seating parts and their covering; replace when
the foam has ceased to provide any protection
Inspect the magnetic lock and cylinder
Inspect all the chair nuts for tightness
Inspect and test the four shock absorbers
Inspect and test the electromagnetic lock
Inspect and test the two manual locks
Inspect and test the harness and the harness buckle
Inspect and test the leg strap
Inspect and test the cylinder and their fastenings
Inspect and test the two axles of rotation and their bearings
Every three years:
Replace the harness by releasing the four snap hooks one by one. When installing the
new harness, make sure that the snap hooks are fully closed and locked:
Cleaning:
Use a soft cloth with a non-aggressive common cleaning agent
Do not use any solvents or aggressive cleaning liquids
Do not use disinfectant sprays