D-0125111-A – 2020/04
TRV chair – Instructions for Use - EN Page 2
1.5 Notes on safety
Throughout this manual, the following meanings of warnings, cautions and notices apply:
The WARNING label identifies conditions or practices that may present
danger to the patient and/or user.
The CAUTION label identifies conditions or practices that could result in
damage to the equipment.
NOTICE
NOTICE is used to address practices not related to personal injury.
1.6 Safety precautions
Follow Instructions for Use - To ensure correct use of the medical device,
it is essential to read this documentation and all the instructions and labels
carefully and thoroughly.
This device must not be used with patients smaller than 140 cm nor taller
than 195 cm, or with patients that weigh more than 150 kg.
The system should not be used in areas of high humidity.
The system should not be exposed to explosive or flammable gases.
The system should only be used by persons who have been trained in its
use and who are medically qualified in the field of vestibulometry.
In the event of damage to the system or any of its components, it must be
repaired before further use.
The device must be cleaned after each utilization.
After/before every manipulation, the user must clean every part in contact
with the patient (shoulders, shims, seat, headrest, head strap) with the
specified cleaning agent.
Always switch the system off before cleaning.
The device must be powered off after each utilization.
After every manipulation, the user must switch off the device to avoid any
unintended activation of the frame, which could hurt the user and the
patient.