Always use the handles to manipulate the device.
Failure to use the handles during maneuvers may incur pinch / trapping
danger and potential damages.
The device may only be used for patients between 140 and 195 cm.
The device is designed to manipulate patients within a certain range of
size (140 cm < S < 195 cm), in order to ensure safe use.
The device must NOT be used for patients that weigh more than 150 kg.
The device is designed to manipulate patients within a certain range of
weight (W < 150 kg), in order to ensure safe use.
In case of a serious incident, the manufacturer should be notified as well
as the competent authority in the patient’s home country.
No modification of this equipment is allowed without authorization from
Interacoustics.
No parts of the equipment can be serviced or maintained while in use with
the patient.
Risk of dizziness.
As the chair allows for 360° rotation.
The use, sale and distribution of the system may be regulated, so it is
essential to ensure that the device is compliant with any local regulations
before being put to use.
Within the European Union, it is illegal to dispose of electrical and
electronic waste as unsorted municipal waste. Electrical and electronic
waste may contain hazardous substances and must therefore be
disposed of separately. Such products will be marked with the crossed-out
wheelie-bin image shown to the left. User cooperation is important in
order to ensure a high level of reuse and recycling of electrical and
electronic waste. Failure to recycle such waste products in an appropriate
way may endanger the environment and consequently the health of
human beings.
Outside the European Union, local regulations should be followed when
disposing of the product.