R
Ricardo PottsAug 13, 2025
What to do if electric actuator fails to lift when button is pressed on Invacare Reliant 450 RHL450-1?
- JJames HallAug 13, 2025
If the electric actuator on your Invacare Medical Equipment fails to lift when the button is pressed, consider these potential causes: * A loose hand-control or actuator connector. * Low battery charge (charge the batteries as described on page 27, or replace the battery pack). * The RED emergency stop button is pressed IN (rotate it CLOCKWISE until it pops out). * The battery is not properly connected to the control box (reconnect as described on page 27). * Damaged connecting terminals. * The electric actuator may need service, or the load is too high.