Invacare®Robin®
Test
Basic
Standard
ComplianceLevel/
FrequencyRange
Mains
Harmonics
EN61000-
3-2:2006+
A1:2009+
A2:2009
ClassA
Voltage
Fluctuations
EN61000-3-
3:2008
dc%/dmax/d(t)/Pst
Electrostatic
Discharge
EN61000-4-
2:1995
±2/4/6kVContact,
±2/4/8kVAir
RadiatedField
Immunity
EN61000-4-
3:2006
3V/m,80%1kHzAM,
80MHz-2.5GHz
EFT/Bursts
EN61000-4-
4:2004
±2kVL/N/E/LNE
SurgeImmunity
EN61000-4-
5:2006
±0.5/1kVL-N,
±0.5/1/2kVL-E/N-E
ConductedRF
Immunity
EN61000-4-
6:2007
3V,80%1kHzAM,
0.15MHz-80MHz
MagneticField
Immunity
EN61000-4-
8:1993
3A/m,50Hz&60Hz
Dips&
Interruptions
EN61000-4-
11:2004
-95%/-60%/-30%
321586745-C