Technical specification
Response time:
T90 < 2 second
Detectable
Range:
0.1 ppm
0 ppm – 5,000 ppm
Resolution:
+/- 0.1 ppm
Accuracy:
+/- 5% displayed reading +/- one digit
Linearity:
+/- 5% displayed reading +/- one digit
Battery:
Lithium ion:
24 hours
Alkaline (Duracell Procell
MN1500):
8.5 hours
Data log:
Including date / time:
80,000
Alarm visual:
Flashing Red and Amber LED
Alarm audible:
95 dBA @ 30 cm
Flow Rate:
≥ 220 ml/min in Ambient conditions
Temperature:
Operating:
-20 to 60 °C (-4 to 140 °F)
Storage:
-25 to 60 °C (-13 to 140 °F)
Certified to:
-15 to 45 °C (-5 to 140 °F)
Dimensions:
Instrument:
340 x 90 x 60 mm
Weight:
Instrument:
0.720 kg (1.6 lb)
Materials:
Instrument:
Anti-static PC/ABS (Polycarbonate/ Acrylonitrile Butadiene Styrene)
Rubber Boot:
Anti-static TPE (Thermoplastic Polyolefin
Elastomeric)